High-density graphite heating module is a product designed and created for samples that is difficult to digest. Its high temperature (the designed temperature may reach 550 ), good uniformity and small
Technical Specifications
| Parameter | Value |
|---|---|
| Repeatability (RSD) | < 0.5% (1~240mgN); calculation and titration process in real time. |
| Recovery Rate | > 99.5% (1~240mgN) |
| Measured Sample Amount | Solid <5g, liquid <20mL |
| Distillation Capacity | 15~30mL/min, adjustable; ensures that the acid-base reaction is controllable to avoid the escape of ammonia caused by |
| Delay Time | 0-1800s (Suitable for measuring nitrate nitrogen) |
| Staged Alkali Addition | Interval 0-9s |
| Printer | Ribbon printer, permanent data storage without color fading; constant steam, uninterrupted steam during water replenishment to ensure smooth vaporization of |
| Digestion Tube Dimension | 300mL |
| External Dimension | 500*350*720mm; -1033- |
| Kjeldahl System | Graphite Digester |